EasyManua.ls Logo

Intercomp CS3000 - Power and Sealing Procedures

Intercomp CS3000
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CS3000, Users
Rev A, January, 2011
Page 25 of 35
Power/Batteries
Remove the two caps in the back of the unit. Tip the old cells out. Change the
cells with standard alkaline “D” batteries, being careful to put the positive end in
first (The end with the bump). Replace the battery caps. Note that the CS3000
will run off a single battery tube, but the battery life will be cut in half.
Legal-for-Trade Sealing
For qualified calibration facilities:
1. After calibration push the calibration switch. The calibration switch is located on the
upper right hand corner on the backside of the scale.
2. Verify that the scale returns to normal weighing when trying to access the weight
calibration.
3. Replace the set screw that covers the cal switch.
4. Apply the protection sticker (sticker must conform to Intercomp’s specifications).

Related product manuals