Note: Number entry: The flashing digit may be incremented by pressing the Lb/Kg
button or decremented by pressing the ZERO button. To move one digit to the left,
press the TARE button. When you have finished entering a number, press the Mode
button. The settings are saved once Mode is pressed to advance the menu, and the
scale can then be turned off.
Setting the Calibration Parameters
1. At any point in the following steps, data will be retained by the scale at the step
completed if the power is cycled off. To initiate calibration press the Mode button.
The scale shows “StEP”. Press the Mode button. The scale shows “000” with the far
right number flashing. To go through all of the calibration parameters, press Mode
button with the display showing “000”. To skip to the weight calibration parameters
and proceed to step 8 enter “002”. See “Troubleshooting” section for additional
codes available. The flashing digit may be incremented by pressing the Lb/Kg button
or decremented by pressing the ZERO button. To move one digit to the left, press
the TARE button. When the desired number is displayed, press the Mode button
2. The display shows “U. EnA” . Press the mode button. The display will read “YES”.
Pressing Lb/Kg or the ZERO button will toggle the display to “no”, which disables the
lb/kg units switching. With the display showing “YES”, press the mode button.
3. The display now shows “-det-”. Press the mode button. The display will read the
current setting. When motion detect is enabled, any motion of the load will disable
the zero, tare, print, and accum functions. Pressing the Lb/Kg or the ZERO button
will toggle the display. When the desired setting is displayed, press the Mode button
4. The display shows “AdC.rt”. Press the mode button. The display shows the current
setting flashing. This sets the internal A/D conversion time at one of two choices.
An entry of ‘0’ results in the full conversion time for the most stable results. An
entry of ‘1’ results in a reduced conversion time which extends battery life. It is
recommended to leave this set to ‘0’. Note that if this setting is changed, the scale
must be recalibrated. The flashing digit may be incremented by pressing the Lb/Kg
button or decremented by pressing the ZERO button. When the desired ADC rate
is displayed press the Mode button.
5. The display shows “AZt”. Press the mode button. The display shows the current
setting. Press the Lb/Kg or the ZERO button to cycle through the auto zero tracking
options (1 d, 3 d, 0.5 d. oFF, or 0.6 d). If the displayed weight is less than the
number of grads shown for a given amount of time, the weight will be zeroed off.
When the desired auto zero tracking setting is displayed press the Mode button.
6. The display shows “ZEr0.r”. Press the mode button. The display shows the current
setting flashing. The flashing digit may be incremented by pressing the Lb/Kg
button or decremented by pressing the ZERO button. The zero range is the
percentage the zero can move from the original zero obtained at calibration. The
zero button will not work if outside the zero range; and the display will show “ZEr0.r