MONITOR TESTING
MONITOR TESTING
Press the TEST key to test the internal functions of the monitor. You should do this each time you begin monitoring
a patient.
The TEST function generates a 1 mV pulse at 70 bpm, causing a waveform and a 70 bpm indication on the display
and a signal at the rear panel connector. If these indications are not present, contact qualified service personnel.
Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments
should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance
with local or governmental regulations. In the event that internal adjustment or recalibration is necessary, refer to
the Service Manual for this equipment.
Model 101R/NR Service Manual 23