2 Reprocessing steps in accordance with ISO 17664 | 2.9 Automatic cleaning programmes
30 / 56
4 Grasp the three-function or multifunctional handpiece on the adapter
between hose and handpiece and take off the cannula ② and the gripping
sleeve ①.
4 Place the three-function or multifunctional handpieces without cannulas ②
or gripping sleeves ① in the middle holders of the hygiene centre.
NOTE
Make sure to place three-function handpieces in the grey holder and multi-
functional handpieces in the black holder of the hygiene centre.
4 Place the instrument hoses of the dentist element in the designated places
of the hygiene centre.
4 Place the suction hoses of the assistant element in the designated places of
the hygiene centre.
① Spray mist suction ④ Three-function handpiece (grey
adapter) or multifunctional hand-
piece (black adapter)
② Saliva ejector ⑤ With insert: Motor / MULTIflex
coupling
③ Without insert: PiezoLED
NOTE
All water regulation mechanisms on the motor, instruments, and MULTIflex
couplings must be open.