Care instructions KaVo uniQa
3 Replacing and checking cleaning and germ reduction agents | 3.4 Checking the OXYGENAL 6 concentration
55 / 56
Other concentrations can be tested using the following table:
Concentration
H
2
O
2
Measuring cup Ratio water :
treated water
Value shown on
test strip
0.02 % (200 ppm)
standard concen-
tration
40ml 39:1 5 mg/l
0.25 % (2500
ppm) intensive
germ reduction
concentration
100ml 99:1 25 mg/l