6-8 External Triggering (Model 2306-VS Only)
Once a channel is enabled, a settings conflict error message will occur when:
1. Trying to change a trigger external setting while
a channel's trigger external enable is
ON. Error message: “+234, Trigger external setting channel enabled conflict.
Error +234 will be generated if BOTH = VOLT or AUTO for channel having
TRIG:EXT
:ENAB OFF command or the other channel has BOTH set to VOLT/AUTO
with external trigger enabled.
2. Trying to change a system parameter that would interfere with a channel’s having its
trig
ger external enable ON. Error message: “+235, Trigger external enabled system
setting conflict.” Error +235 will be generated if external trigger is enabled and these
commands are sent: *TRG, *TRG1, *TRG2, READ1?, READ2?, MEAS?,
BOTHREAD?, BOTHFETCh?, and BOTHTRG.
TRIGger[1]:EXTernal:ENABle:INIT Applies to battery channel (#1)
TRIGger2:EXTernal:ENABle:INIT Applies to charger channel (#2)
These commands initialize the channel to take another set
of STEP:POINts measurements.
For configurations taking measurements, the measurements will stop after taking the POINts
number of measurements. This allows time for sending FETCh:ARRay? query command to get
the data from the buffer before being overwritten by another measurement. However, if voltage
stepping only and not taking measurements, the list will cycle around to step 1 and continue
stepping voltages and not stop. After querying for measurements, send this command to
initialize the channel to resume detecting and responding to trigger in pulses on the channel.
This command must be sent while the trigger enable is set to ON. If not sent, doing a
FETCh:ARRay? query will result in the same data being returned. Send the INIT command if:
1. A channel is taking AUTO measurements (READ = AUTO).
2. A channel is taking measurements and volt stepping (READ = AUTO or SYNC; VOLT
ON).
3. A channel is taking AUTO measurements on both channels (BOTH = AUTO).
After processing the INIT command and ready to detect
trigger in pulses again, the trigger
out signal will pulse accordingly to indicate it is ready for the next trigger in. This command will
generate a settings conflict error message if the INIT command is sent when the channel's trigger
external functionality is disabled, or if the INIT command is being requested while the channel’s
trigger external BOTH setting is NONE and the other channel’s trigger external BOTH setting
is AUTO or VOLT. Error message: “+224, Trigger external enable init conflict.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals