6-16 External Triggering (Model 2306-VS Only)
Programming examples
Example 1. Measure the high current in a chain of GSM signal with triggering.
Model 2306-VS configuration using channel 1:
*RST ‘ reset the unit
:VOLT 6 ‘ set voltage
:CURR 2.25 ‘ set current
:SENS:PCUR:FAST ON ‘ turn fast mode on
:SENS:PCUR:TIME:HIGH 500e-6 ‘ set high current time (this must be set
‘ in external triggering mode because
‘ no AUTO detection exists)
:SENS:PCUR:MODE HIGH ‘ measure high current
:SENS:FUNC 'PCUR' ‘ switch to pulse current functionality
:TRIG:EXT:STEP:VOLT OFF ‘ turn off the voltage step
:TRIG:EXT:STEP:READ AUTO ‘ switch read mode to AUTO, means
‘ 2306-VS will measure right after it is
‘ triggered
:TRIG:EXT:STEP:POIN 8 ‘ set the number of points to be
‘ triggered and put into the buffer
Loop 1:
:TRIG:EXT:ENAB ON ‘ turn external triggering ON. 2306-VS
‘ is waiting for the trigger signal.
Loop 2:
Monitor for trigger out pulse on channel 1
Generate 8 trigger in pulses on channel 1 and monitor for trigger out pulses
on channel 1 - make sure a new trigger in pulse occurs after each trigger out
pulse. Trigger in pulses are ignored until the trigger out pulses. Trigger
ins detected on falling edge and trigger outs pulse low for about 10us.
:FETCH:ARRAY? ‘ get the data back
:TRIG:EXT:ENAB:INIT ‘ init external trigger to take
‘ another set of 8 measurements
go back to Loop 2
:TRIG:EXT:ENAB OFF ‘ when done taking
‘ measurements or desire to make
‘ changes
go back to Loop 1 after making changes
NOTE If you are unable to monitor for trigger out pulses, either:
(1) monitor the buffer full bit in the status model (Section 8), or
(2) allow sufficient time between trigger in pulses to allow the unit to complete each
step (see the specifications in Appendix A for time guidelines).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals