C-6 Calibration Reference
Detecting calibration errors
If an error occurs during any calibration step, the Model 2306 will generate an appropriate
error message. Several methods to detect calibration errors are discussed below.
Reading the error queue
As with other Model 2306 errors, any calibration errors will be reported in the error queue.
Use the :SYST:ERR? query to read the error queue.
Error summary
Table C-2 summarizes calibration errors.
Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it
is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query to
read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an
error occurs, send the following command: *SRE 4. This command will enable SRQ when the
EAV bit is set. You can then read the status byte and error queue as outlined above to check for
errors and to determine the exact nature of the error.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals