Calibration 14-11
Step 4: Enter calibration dates, and save calibration
1. After completing all Channel #1 and Channel #2 calibration steps, the unit will prompt
if you wish to save calibration:
CALIBRATE UNIT
Save Cal Data YES
2. To save new calibration constants, select YES, then press ENTER. If you wish to exit
calibration without saving new calibration constants, select NO, then press ENTER. In
that case, the unit will revert to prior calibration constants.
3. The unit will then prompt you to enter the calibration date:
CALIBRATE UNIT
Cal Date 02/01/2003
4. Using the edit keys, set the calibration date to today’s date, then press ENTER. The unit
will display the following:
CALIBRATE UNIT
EXITING CAL
5. Press ENTER to complete the calibration procedure and return to the menu display.
Calibration is now complete.
Remote calibration
Follow the procedure outlined below to perform Model 2306 remote calibration by sending
SCPI commands over the IEEE-488 bus. The remote commands and appropriate parameters are
separately summarized for each step.
Remote calibration display
The Model 2306 will display the following while being calibrated over the bus:
CALIBRATING UNIT
FROM THE BUS R
NOTE The “R” located on the lower right hand corner of the display indicates remote
operation.