EasyManua.ls Logo

Keithley 2302 - Trigger on Talk both Channels; Bus Commands

Keithley 2302
348 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
488.1 Protocol G-5
Trigger on talk both channels
NOTE This command applies only to Models 2306, 2306-VS and 2306-PJ.
Trigger on talk will return a reading for the active channel (DISPlay:CHANnel?) unless you
ha
ve enabled (ON) the setting for Trig on talk to return two readings—one from each channel
comma delimited (for example: ch1 reading, ch2 reading). This command maps trigger on talk
to BOTHREAD? when enabled.
To enable the two readings trig on talk and have it apply to both channels (READ1? for
chan
nel 1, READ 2? for channel 2), send:
SYSTem:TRIGger:TALK:BOTH ON
To disable the two readings trig on talk, send:
SYSTem:TRIGger:TALK:BOTH OFF
Bus commands
Table G-1 contains bus commands for trigger on talk,
Table G-1
Trigger on talk bus commands
Commands Description Default
:SYSTem
:TRIGger
:TALK
:BOTH <b>
:BOTH?
Path to System commands.
Path to trigger commands.
Path to talk commands.
Set <b> to 1 (or ON) to enable 2 reading trig on talk. 0 (or OFF)
disable 2 re
ading trig on talk, but reads the active channel:
CH1: READ1?
CH2: READ2?
Query trigger on talk.
OFF