Pulse Current Measurements 3-5
sure to factor in the trigger delay (both internal plus user) when determining integration times
(see Figure 3-2). When manually set using the front panel keys, the values are changed in
increments of 33.3333μsec. Th
is ensures that an integral value of 33.3333μsec will be selected.
NOTE A
uto time does not account for user trigger delay — if using auto time, make sure the
user trigger delay is appropriately set for the desired overall measurement time.
NOTE Do not u
se auto time with the Model 2306-VS if the other channel has trigger external
enabled (Section 6).
After auto time acquires a time value (auto time), the auto time is adjusted for the internal
tri
gger delay of 15μsec (auto internal time). The auto internal time is then adjusted to be an
integral time value of 33.3333μsec (auto integral time). For example:
auto time value = 28.053msec
auto internal time = 28.053ms - 0.015msec = 28.038ms
auto integral time = 28.033ms (response returned when time setting is queried)
When a pulse time is set via the bus, the time is assumed to be an auto internal time (i
.e., the
value is assumed to be adjusted for the internal delay value). This value is then adjusted to the
applicable integral value. For example:
manual time value = 5.040msec
integral time = 5.033ms (response returned when time setting is queried)
Average readings count
NOTE The menu item AVER READINGS #1/#2 applies to average readings for DVM, I and
V where AVERAGE READINGS under PULSE CURRENT #1/#2 menu item applies
to pulse current measurements.
The average readings count specifies how many measurements (integrations) are performed
and a
veraged for each displayed reading. For example, assume that the pulse average readings
count is 10 and you are measuring PULSE HIGH. Each displayed reading will reflect the
average of 10 peak pulse measurements.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com

Table of Contents

Other manuals for Keithley 2302

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2302 and is the answer not in the manual?

Keithley 2302 Specifications

General IconGeneral
BrandKeithley
Model2302
CategoryPower Supply
LanguageEnglish

Related product manuals