12-8 SCPI Tables
:SENSe[1]
:PCURrent
:STEP
:DELay
<NRf> Step delay: 0msec–100msec (in 10μsec steps). 0
:SEQuence Define up to 19 delay sequence values using the index
and the time v
alue (e.g. "1,1e-3" sets the first delay
to 1ms). Ranges: i: 1-19 and t: 0 to 100e-3.
:LENgth Define the delay sequence length (0 to 19). Specify
zero
to always use the PCUR:STEP:DELay value.
0
:SKIP
Specify whether (1) or not (0) to skip sequential
trig
gers.
1=Only requires 1 trigger
0 = Requires multiple triggers
0
:DELay? Query setting for step delay.
:RANGe
<NRf> Set step trigger level range for 5A current range:
100mA, 1A, 5A.
5A
:RANGe? Query setting for step trigger level range for 5A
curr
ent range.
:MILLiamp
<NRf>
Model 2306-PJ only — when on 500mA current
range.
The parameter <NRf> sent with this
command causes the trigger to be set with the
trigger level setting of HALFamp, HUNDred, or
TEN.
:MILLiamp? Model 2306-PJ only — Query the trigger level
se
tting on the 500mA current range. Receives
responses in amps of 0.5, 0.1, or 0.01,
accordingly.
:TLEVx
<NRf> Set trigger level for each TLEV step (x equals 1–20)
(0.0–maxA where max is 100mA for 100mA
range, 1A for 1A range, and 5A for 5A range
setting).
0.0
:TLEVx? Query setting for trigger level for each TLEV step
(x equals 1–20).
:SYNChronize Pulse detection triggering:
[:STATe] <b> Send ON to select pulse current measurements or OFF
to select pulse
current digitization.
ON
Table 12-4 (cont.)
SENSe command summary (refer to Tables 2-3, 3-2, and 4-2) (cont.)
Command Description
Default
parameter
SCPI