12-20 SCPI Tables
Table 12-8
Model 2306-VS external trigger command summary (refer to Section 6)
Command Description Default SCPI
TRIGger[1] Trigger subsystem for Channel #1 (battery channel):
:EXTernal Path for external trigger control.
:BOTH <name> Select channel both control type (NONE,
VO
LT, or AUTO).
NONE
:BOTH? Query channel both control type.
:EDGE Path for specifying the trigger in & out pulse
edge
s for a channel.
:IN <name> Select trigger in pulse polarity (RISING or
F
ALLING).
FALLING
:IN? Query trigger in pulse polarity.
:OUT <name> Select trigger out pulse polarity (RISING or
F
ALLING).
FALLING
:OUT? Query trigger out pulse polarity.
:ENABle <b> Enable (ON) or disable (OFF) external trigger
feature.
OFF
:ENABle? Query external trigger feature state.
:INIT Initialize channel to take another set of
STEP:POINts measurements.
:STEP
<NRf1
>,<NRf2>,<NRf3>
Set voltage step parameters; step # (1-20),
voltage (0-15), delay (0-5).
Step 1-
20,
0V step,
0s delay
:STEP? <NRf> Query voltage and delay for a given step
number
.
:STEP? Query next step to be completed upon detection
of the
trigger in pulse.
:VOLTage <b> Enable (ON) or disable (OFF) voltage step
feature.
OFF
:VOLTage? Query voltage step state.
:END <NRf> Set voltage value when external trigger
turned OFF
(0-15).
0
:END? Query trigger OFF voltage.
:READing <name> Set reading/measurement mode (NONE,
SYNC or
AUTO).
AUTO
:READing? Query reading/measurement mode.
:POINts <NRf> Set # of measurements and/or voltage steps
(1-
20).
1
:POINts? Query # of measurements and/or voltage
ste
ps.
:VPT <b> Enable (ON) or disable (OFF) Voltage
Protection T
ripping.
ON