7-18 Keysight N5221A/22A Service Guide
Repair and Replacement Procedures
Removing and Replacing the A14 System Motherboard
7-
Removing and Replacing the A14 System Motherboard
Tools Required
T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
T-20 TORX driver (torque to 21 in-lb or 2.38N.m)
5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
ESD grounding wrist strap
Removal Procedure
Refer to Figure 7-7 for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to “Removing the Covers” on
page 7-6.
3. Remove the front panel assembly. Refer to “Removing and Replacing the
Front Panel Assembly” on page 7-8.
4. Remove the A18 GPIB board. Refer to “Removing and Replacing the A18
GPIB Board” on page 7-27.
5. Remove the A15 midplane board. Refer to “Removing and Replacing the
A15 Midplane Board” on page 7-21.
6. Remove the A4–A13 boards. Refer to “Removing and Replacing the A4–
A13 Boards” on page 7-15.
7. Remove the fan bracket and fans. Refer to “Removing and Replacing the
Fans” on page 7-56.
8. Turn the analyzer over so that the bottom side is up and remove four
screws (item
), that secure the bottom of the midplane bracket to the
chassis.
9. Turn the analyzer back over so that the top side is up. Remove seven
screws (item
) that secure the left side bracket to the chassis left inner
panel.
10.Remove two screws (item
) from each side that secure the midplane
bracket to the chassis left and right inner panels. Lift the midplane bracket
out of the analyzer.
11.Remove three screws (item
) that secure the A14 system motherboard
to the chassis.
12.Slide the A14 system motherboard toward the rear of the analyzer to
release it from the 15 keyhole standoffs (item
) on the chassis.
13.Lift the A14 system motherboard just enough to unplug the ribbon cable
from the bottom side, then lift it out of the analyzer.