2-4 Keysight N5221A/22A Service Guide
General Product Information
Required Service Test Equipment
2-
Required Service Test Equipment
Equipment
a
a. Unless specified otherwise, equipment listed is required for all analyzer models.
Critical Specifications Recommended
Model or Part
Number
Alternate Model or
Part Number
Use
b
b. P = Performance tests, A = Adjustments, T = Troubleshooting, V = System verification
Test Instruments and Software
Compression
test set
None specified U3070AK01 None P
Dynamic accuracy
test set
None specified U3020AD01 None P
Test software
c
c. The recommended model or part number for all equipment listed with a “P” in the Use column is required for proper operation
of this test software.
N/A N7840A None P
Frequency counter Freq: 10 MHz to 20 GHz
Accuracy : ±0.5 ppm
53151A
Opt 001
None P, A,T
Signal generator CW Freq: 1.185 GHz N5181A, Option 503 E8257D, Option 520 P, A, T
Power meter Accuracy: ±0.0068 dB N1913A/14A E4418A/B,
E4419A/B
d
d. If an accurate measurement of the dynamic accuracy specification is not required, the E4418A or E4419A can be used.
P, A , T
Power sensor Freq: 10 MHz to 4.2 GHz
Range: –30 to +20 dBm
N8482A 8482A P, A,T
Power sensor Freq: 50 MHz to 40 or 50 GHz
Range: –30 to +20 dBm
N8485A 8485A or U8485A P, A,T
Spectrum analyzer Min Freq: 1 MHz
Max Freq: > 4 GHz
Resolution BW: 300 Hz
E444xA PSA series,
N90xxA signal
analyzer family
856xE A,T
Digital voltmeter Resolution: 10 mV Any Any T
Printer N/A Any printer with Microsoft Windows XP or
Windows 7 driver
--
Mouse N/A Any Any --
Keyboard N/A Any Any --