CALibrate:CURRent:MEASure <value>
Calibrates the low range current measurement. The value selects the range to calibrate.
Parameter Typical Return
The maximum current of the measurement range. (none)
Calibrates the current of the 10 A measurement range:
CAL:CURR:MEAS 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter Typical Return
(none) (none)
Calibrates the current sharing: CAL:CURR:SHAR
CALibrate:CURRent:TC
Calibrates the temperature coefficient.
Parameter Typical Return
(none) (none)
Calibrates the temperature coefficient: CAL:CURR:TC
CALibrate:DATA <value>
Enters the calibration value read by the external meter. You must first select a calibration level for the
value being entered. Data values are expressed in base units - either volts or amperes, depending on
which function is being calibrated.
Parameter Typical Return
Numeric value (none)
Specify calibration value 0.0237: CAL:DATA 2.37E-2
CALibrate:DATE <"date">
CALibrate:DATE?
Enters the calibration date in nonvolatile memory. Enter any ASCII string up to 15 characters. The
query returns the date. Enclose string parameters in single or double quotes.
6 SCPI Programming Reference
215 Keysight N6900/N7900 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N6972A with Option 301 and is the answer not in the manual?

Keysight Technologies N6972A with Option 301 Specifications

General IconGeneral
BrandKeysight Technologies
ModelN6972A with Option 301
CategoryPower Supply
LanguageEnglish

Related product manuals