CALibrate:CURRent:MEASure <value>
Calibrates the low range current measurement. The value selects the range to calibrate.
Parameter Typical Return
The maximum current of the measurement range. (none)
Calibrates the current of the 10 A measurement range:
CAL:CURR:MEAS 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter Typical Return
(none) (none)
Calibrates the current sharing: CAL:CURR:SHAR
CALibrate:CURRent:TC
Calibrates the temperature coefficient.
Parameter Typical Return
(none) (none)
Calibrates the temperature coefficient: CAL:CURR:TC
CALibrate:DATA <value>
Enters the calibration value read by the external meter. You must first select a calibration level for the
value being entered. Data values are expressed in base units - either volts or amperes, depending on
which function is being calibrated.
Parameter Typical Return
Numeric value (none)
Specify calibration value 0.0237: CAL:DATA 2.37E-2
CALibrate:DATE <"date">
CALibrate:DATE?
Enters the calibration date in nonvolatile memory. Enter any ASCII string up to 15 characters. The
query returns the date. Enclose string parameters in single or double quotes.
6 SCPI Programming Reference
215 Keysight N6900/N7900 Series Operating and Service Guide