Parameter Typical Return
(none) 0 (pass) or +1 (failed)
Perform self-test: *TST?
l The power-on self-test is the same self-test performed by *TST.
l *TST? also forces an *RST command.
*WAI
Pauses additional command processing until all pending operations are complete. See OPC for more
information.
Parameter Typical Return
(none) (none)
Waituntil all pending operations complete. *WAI
l *WAI can only be aborted by sending the instrument a Device Clear command.
Keysight N6900/N7900 Series Operating and Service Guide 238
6 SCPI Programming Reference