E8257D/67D, E8663D PSG Signal Generators Service Guide
Troubleshooting
Troubleshooting Unlevels
1-68
d. If any of the frequency ranges do not produce power levels the maximum leveled power shown in the tables
below, set the signal generator to the frequency with the lowest power level and measure the power with a
power meter. If the power level is low, troubleshoot the RF path before proceeding.
Table 1-17 E8257D Maximum Leveled Power (dBm)
a
a
always check data sheet for updates to specifications
Frequency 20 GHz Models 40 GHz Models 50 & 67 GHz Models
Std. 1EA/
1EU
1E1 1E1+
1EA/
1EU
Std. 1EA/
1EU
1E1 1E1+
1EA/
1EU
Std. 1EA/
1EU
1E1 1E1+
1EA/
1EU
250 kHz to 3.2 GHz +13 +16 +11 +15 +9 +15 +7 +14 +5 +14 +3 +13
250 kHz to 3.2 GHz w/UNW +11 +11 +10 +10 +9 +10 +7 +9 +5 +9 +3 +8
250 kHz to 3.2 GHz w/1EH
b
b
with filters switched out
+13 +13 +11 +12 +9 +12 +7 +11 +5 +11 +3 +10
250 kHz to 3.2 GHz
w/UNW + 1EH
b
+10 +10 +9 +9 +9 +9 +7 +8 +5 +8 +3 +7
>3.2 GHz to 20 GHz +13 +20 +11 +18 +9 +18 +7 +16 +5 +14 +3 +13
>20 GHz to 30 GHz +9 +14 +7 +12 +5 +11 +3 +9
>30 GHz to 40 GHz +9 +14 +7 +12
>30 GHz to 65 GHz +5 +11 +3 +9
>65 GHz to 67 GHz +5 +10 +3 +8
>67 GHz to 70 GHz +5 +8 +3 +6
Table 1-18 E8267D Maximum Leveled Power (dBm)
a
a
Always check data sheet for updates to specifications
Frequency 20 GHz Models 32 & 44 GHz Models
250 kHz to 3.2 GHz +13 +12
250 kHz to 3.2 GHz w/UNW +9 +8
250 kHz to 3.2 GHz w/1EH
b
b
With filters switched out
+10 +9
250 kHz to 3.2 GHz w/UNW + 1EH
b
+7 +6
>3.2 GHz to 20 GHz +18 +14
>20 GHz to 31.8 GHz +14
>31.8 GHz to 40 GHz +12
>40 GHz to 44 GHz +10

Table of Contents