68 Keysight InfiniiVision 1000 X-Series Oscilloscopes User's Guide
2 Quick Reference
In a mask file, all region definitions need to be separated by a blank line.
Mask regions are defined by a number of (x,y) coordinate vertices (as on an
ordinary x,y graph). A "y" value of "MAX" specifies the top of the graticule, and a
"y" value of "MIN" specifies the bottom of the graticule.
The mask x,y graph is related to the oscilloscope graticule using the
:MTESt:SCALe setup commands.
The oscilloscope's graticule has a time reference location (at the left, center, or
right of the screen) and a trigger (t=0) position/delay value relative to the
reference. The graticule also has a vertical ground 0 V reference (offset relative to
the center of the screen) location.
The X1 and Y1 setup commands relate the mask region's x,y graph origin to the
oscilloscope graticule's t=0 and V=0 reference locations, and the XDELta and Y2
setup commands specify the size of the graph's x and y units.
The X1 setup command specifies the time location of the x,y graph's x origin.
The Y1 setup command specifies the vertical location of the x,y graph's y origin.
The XDELta setup command specifies the amount of time associated with each
x unit.
The Y2 setup command is the vertical location of the x,y graph's y=1 value (so
in effect, Y2 – Y1 is the YDELta value).
For example:
With a graticule whose trigger position is 10 ns (before a center screen
reference) and whose ground reference (offset) is 2 V below the center of the
screen, to place the mask region's x,y graph's origin at center screen, you would
set X1 = 10 ns and Y1 = 2 V.
If the XDELta parameter is set to 5 ns and Y2 is set to 4 V, a mask region whose
vertices are (-1, 1), (1, 1), (1, -1), and (-1, -1) goes from 5 ns to 15 ns and from
0V to 4V.
If you move the mask region's x,y graph origin to the t=0 and V=0 location by
setting X1 = 0 and Y1 = 0, the same vertices define a region that goes from
-5 ns to 5 ns and from -2 V to 2 V.
NOTE
Although a mask can have up to 8 regions, in any given vertical column, it is only possible to
define 4 regions. When there are 4 regions in a vertical column, one region must be tied to the
top (using the MAX y value) and one must be tied to the bottom (using the MIN y value).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight InfiniiVision DSOX1102G and is the answer not in the manual?

Keysight InfiniiVision DSOX1102G Specifications

General IconGeneral
BrandKeysight
ModelInfiniiVision DSOX1102G
CategoryTest Equipment
LanguageEnglish

Related product manuals