How is mask testing done? InfiniiVision oscilloscopes start mask testing by
creating a database that is 200 x 640 for the waveform viewing area. Each location
in the array is designated as either a violation or a pass area. Each time a data
point from a waveform occurs in a violation area a failure is logged. If Test All was
selected, every active analog channel is tested against the mask database for
each acquisition. Over 2 billion failures can be logged per-channel. The number of
acquisitions tested is also logged and displayed as "# of Tests".
The mask file allows greater resolution than the 200 X 640 database. Some
quantization of data occurs to reduce the mask file data for display on-screen.