Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide 125
Remote Programming 4
:INSTrument Commands
:INSTrument:DACMode[?]
Command :INSTrument:DACMode SINGle|DUAL|FOUR|MARKer|DCMarker
Query :INSTrument:DACMode?
Parameters SINGle|DUAL|FOUR|MARKer|DCMarker
SINGle – Channel 1 can generate a signal
DUAL – Channels 1 and 2 can generate a signal, channels 3 and 4 are
unused
FOUR – Channels 1, 2, 3, and 4 can generate a signal
MARKer – Channel 1 with two markers output on channel 2 and 3
DCMarker – Dual channel with marker: Channels 1 and 2 can generate
a signal. Channel 1 has two markers output on channel 3 and 4.
Channel 2 can generate signals without markers.
Parameter Suffix None
Description This command sets the operation mode of the DAC. The value of the
operation mode determines, to which channels waveforms can be
transferred and the format of the waveform data. In operation mode
SINGle, DUAL, or FOUR the data consists of 1-byte waveform samples
only. In operation mode MARKer or DCMarker the data loaded to channel
1 consists of interleaved 1-byte waveform and 1-byte marker samples
Examples Command
:INST:DACM DUAL
Query
:INST:DACM? -> DUAL
:INSTrument:HWRevision?
Command :INSTrument:HWRevision?
Parameters None
Parameter Suffix None
Description This query returns the M8194A hardware revision number.