162 Keysight M8194A 120 GSa/s Arbitrary Waveform Generator User Guide
6 Characteristics
General
WARNING
The instrument is not designed for outdoor use. Do not expose the
instrument to rain or other excessive moisture. Protect the instrument
from humidity and temperature changes, which could cause
condensation within the instrument.
Do not operate the instrument in the presence of flammable gases,
fumes or powders. Operation of any electrical instrument in such an
environment constitutes a definite safety hazard.
Power consumption 50 W (nom)
Safety designed to IEC61010-1, UL61010, CSA22.2 61010.1 tested
EMC tested to IEC61326-1
Warm-up time 30 min
Calibration interval 2 years recommended
Cooling
Requirements
When operating the M8194A choose a location that provides at least 80 mm of
clearance at rear, and at least 30 mm of clearance at each side for the AXIe
chassis.