Power measurements require simultaneous measurement capability (refer to Model Differences).
If a FETCh query is sent before the measurement is finished, the response will be delayed until the
measurement trigger occurs and the acquisition completes. You can test the MEAS_active bit in the
operation status register to know when the measurement trigger system has returned to the idle
state.
Front Panel Menu Reference SCPI Command
Not available To query the MEAS_active bit(bit5):
STAT:OPER:COND? (@1)
If a bit value of 32 is returned in the query, the MEAS_active bit is true, and the measurement is NOT
complete. When the MEAS_active bit is false, you can retrieve the measurement. Refer to Status
Tutorial for more information.
Keysight N6700C Operating and Service Guide 143
4 Operating the Power System