[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:TRAPezoid:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:TRAPezoid:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:TRAPezoid:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:TRAPezoid:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:TRAPezoid:TOP:TIMe <value>, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:TRAPezoid:TOP:TIMe? [MIN|MAX,] <step#>, (@<chanlist>)
Specifies the time of the top level of the trapezoid in seconds for Arb sequence steps.
Parameter Typical Return
0 – 262.144 | MIN | MAX, *RST 0 <top time>
Step 0 - 99, *RST 0 (none)
Programs a top time for the current trapezoid: ARB:CURR:TRAP:TOP:TIM 2,(@1)
Programs a top time for the voltage trapezoid: ARB:VOLT:TRAP:TOP:TIM 2,(@1)
Programs a top time for the power trapezoid: ARB:POW:TRAP:TOP:TIM 2,(@1)
Programs a top time for the resistance trapezoid: ARB:RESISTANCE:TRAP:TOP:TIM 2,(@1)
Programs a top time for step 0: ARB:SEQ:STEP:CURR:TRAP:TOP:TIM 2, 0,(@1)
Keysight N6700C Operating and Service Guide 219
5 SCPI Programming Reference