[SOURce:]ARB:CURRent:UDEFined:BOSTep:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:BOSTep:POINts? (@<chanlist>)
[SOURce:]ARB:POWer:UDEFined:BOSTep:POINts? (@<chanlist>)
[SOURce:]ARB:RESISTANCE:UDEFined:BOSTep:POINts? (@<chanlist>)
Returns the number of beginning-of-step points in the user-defined Arb, not the point values.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:BOSTep:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:BOSTep:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:POWer:UDEFined:BOSTep:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:RESISTANCE:UDEFined:BOSTep:POINts? <step#>, (@<chanlist>)
Returns the number of beginning-of-step points in the user-defined Arb for Arb sequence steps.
Parameter Typical Return
(none) <number of points>
Step 0 - 99, *RST 0 (none)
Returns the number of current BOST points: ARB:CURR:UDEF:BOST:POIN? (@1)
Returns the number of voltage BOST points: ARB:VOLT:UDEF:BOST:POIN? (@1)
Returns the number of power BOST points: ARB:POW:UDEF:BOST:POIN? (@1)
Returns the number of resistance BOST points: ARB:RESISTANCE:UDEF:BOST:POIN? (@1)
Returns the number of BOST points for step 0: ARB:SEQ:STEP:CURR:UDEF:BOST:POIN? 0,(@1)
[SOURce:]ARB:CURRent:UDEFined:DWELl <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:CURRent:UDEFined:DWELl? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl? (@<chanlist>)
[SOURce:]ARB:POWer:UDEFined:DWELl <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:POWer:UDEFined:DWELl? (@<chanlist>)
[SOURce:]ARB:RESISTANCE:UDEFined:DWELl <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:RESISTANCE:UDEFined:DWELl? (@<chanlist>)
Specifies the dwell time for each user-defined Arb point. A comma-delimited list of up to 511 points
may be programmed. Dwell time is the time that the output will remain at a specific point. Dwell times
can be programmed from 0 through 262.144 seconds with the resolution shown in the table below:
Keysight N6700C Operating and Service Guide 221