[SOURce:]POWer:MODE FIXed|STEP|LIST|ARB, (@<chanlist>)
[SOURce:]POWer:MODE? (@<chanlist>)
Sets the transient mode. This determines what happens to the input power when the transient system
is initiated and triggered.
FIXed - keeps the input power at its immediate value.
STEP - steps the input power to the triggered level when a trigger occurs.
LIST - causes the input power to follow the list values when a trigger occurs.
ARB causes the input power to follow the arbitrary waveform values when a trigger occurs.
Parameter Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed FIX, STEP, LIST, or ARB
Sets the input power mode to Step: POW:MODE STEP, (@1)
[SOURce:]POWer:PROTection:DELay[:TIME] <value>, (@<chanlist>)
[SOURce:]POWer:PROTection:DELay[:TIME]? [MIN|MAX,] (@<chanlist>)
This command is the same as OUTPut:PROTection:DELay.
Sets the over-power protection delay. The over-power protection function will not be triggered during
the delay time. This prevents momentary changes in output status from triggering the over-power
protection function. If the input is regulating in power limit longer than the programmable delay time,
the OPP will trip, and the input will be turned off. Programmed values can range from 0 to 0.255
seconds with a resolution of 1 millisecond.
Parameter Typical Return
0 ā 0.255|MIN|MAX, *RST 0.020 <delay value>
Sets the protection delay to 0.2 seconds: POW:PROT:DEL 0.2, (@1)
[SOURce:]POWer:PROTection:STATe 0|OFF|1|ON, (@<chanlist>)
[SOURce:]POWer:PROTection:STATe?[MIN|MAX,] (@<chanlist>)
Enables or disables the power protection. When the input power exceeds the power rating of the
module by 10%, the output is disabled and the Questionable Condition status register CP+ bit is set.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0īorī1
Enable the power protection state: POW:PROT:STATON, (@1)
l A power protection condition can be cleared with OUTPut:PROTection:CLEar after the cause of the
condition is removed.
Keysight N6700C Operating and Service Guide 273
5 SCPI Programming Reference