Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable current measurement autoranging: SENS:CURR:RANG:AUTO ON, (@1)
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer] <value>, (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe[:UPPer]? [MIN|MAX,] (@<chanlist>)
Selects the Elog current or voltage measurement range. The value that you enter must be the highest
value that you expect to measure. Units are in amperes or volts. The instrument selects the range
with the best resolution for the value entered.
Parameter Typical Return
0 to maximum|MIN|MAX <maximum value of range>
Selects the 1 A range: SENS:ELOG:CURR:RANG 1, (@1)
Selects the 2 V range: SENS:ELOG:VOLT:RANG 2, (@1)
l The query returns the maximum current that can be measured on the range that is presently set.
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:CURRent[:DC]:RANGe:AUTO? (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO 0|OFF|1|ON , (@<chanlist>)
SENSe:ELOG:VOLTage[:DC]:RANGe:AUTO? (@<chanlist>)
Enables or disables seamless measurement autoranging.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable Elog measurement autoranging:
SENS:ELOG:CURR:RANG:AUTO ON, (@1)
SENS:ELOG:VOLT:RANG:AUTO ON, (@1)
SENSe:ELOG:FUNCtion:CURRent 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage? (@<chanlist>)
Enables or disables the Elog current or voltage measurement function.