Parameter Typical Return
0|OFF|1|ON,
*RST current OFF; voltage ON
0îorî1
Enables datalog current measurements: SENS:ELOG:FUNC:CURR ON, (@1)
Enables datalog voltage measurements: SENS:ELOG:FUNC:VOLTON, (@1)
SENSe:ELOG:FUNCtion:CURRent:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:CURRent:MINMax? (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax 0|OFF|1|ON, (@<chanlist>)
SENSe:ELOG:FUNCtion:VOLTage:MINMax? (@<chanlist>)
Enables or disables logging of the minimum and maximum current or voltage values.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0îorî1
Enables MIN/MAX current logging values: SENS:ELOG:FUNC:CURR:MINM ON, (@1)
Enables MIN/MAX voltage logging values: SENS:ELOG:FUNC:VOLT:MINM ON, (@1)
SENSe:ELOG:PERiod <value>, (@<chanlist>)
SENSe:ELOG:PERiod? [MIN|MAX,] (@<chanlist>)
Sets the integration time of an Elog measurement. Although the absolute minimum logging period is
102.4 microseconds, the actual minimum varies as a function of the number of readings that are being
logged (see Integration Period).
Parameter Typical Return
0.0001024 to 60|MIN|MAX, *RST MAX <period>
Specifies a datalog period of 0.01 seconds: SENS:ELOG:PER 0.01, (@1)
SENSe:FUNCtion "CURRent"|"VOLTage"|"NONe", (@<chanlist>)
SENSe:FUNCtion? (@<chanlist>)
Selects a measurement function on models that do not have simultaneous voltage and current
measurement capability. This command is required so that the acquisition system knows which
function to acquire when a measurement is triggered. Selecting NONE disables the measurement
function.
Parameter Typical Return
"CURRent"|"VOLTage"|"NONE", *RST VOLTage CURR, VOLT, or NONE
Selects the current measurement function: SENS:FUNC "CURR", (@1)
5 SCPI Programming Reference
282 Keysight N6700C Operating and Service Guide