Low Range Voltage Programming - Models N6781A, N6782A, N6784A
Step 1. Select the low-voltage programming range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 6 V range of output 1.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Volt\Vprog.
Select the Low1 range. Then press Next.
CAL:VOLT 6, (@1)
Step 2. Select the first voltage calibration point and enter the data. This value may be positive or
negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external DMM.
Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Step 4 applies only to models N6781A and N6782A
Step 4. Repeat steps 1 through 3 for the 600 mV range. This time, select the Low2 range.
Steps 5 and 6 apply only to model N6784A
Step 5. Select the third voltage calibration point for the negative range and enter the data. This value
should be negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P3 measured data”
Enter the data from the external DMM. Press Select when done. Press Next to
finish calibration.
CAL:LEV P3
*OPC?
CAL:DATA <data>
Step 6. Repeat steps 1 through 3, and 5 for the 600 mV range. This time, select the Low2 range.
Low Range Voltage Measurement - Models N6781A, N6782A, N6784A
Step 1. Select the low-voltage measurement range. The value entered must be the maximum voltage
of the range to be calibrated. This example selects the 1 V range of output 1.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Volt\Vmeas.
Select the Low1 range. Then press Next.
CAL:VOLT:MEAS 1, (@1)
Keysight N6700C Operating and Service Guide 361
6 Verification and Calibration