Step 2. Select the first voltage calibration point and enter the data. This value may be positive or
negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data”Enter the data from the external DMM.
Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 3. Select the second voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P2 measured data”Enter the data from the external
DMM.Press Select when done.
Press Next to finish calibration.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Step 4 applies only to models N6781A and N6782A
Step 4. Repeat steps 1 through 3 for the 100 mV range. This time, select the Low2 range.
Steps 5 and 6 apply only to model N6784A
Step 5. Select the third voltage calibration point for the negative range and enter the data. This value
should be negative.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P3 measured data”
Enter the data from the external DMM. Press Select when done. Press Next to
finish calibration.
CAL:LEV P3
*OPC?
CAL:DATA <data>
Step 6. Repeat steps 1 through 3, and 5 for the 100 mV range. This time, select the Low2 range.
Positive Voltage Limit - Models N6781A, N6782A, N6784A
Step 1. Select the positive voltage limit range. The value to program a range must be the maximum
voltage of the range to be calibrated. This example selects the 20 V positive limit range of output 1.
Front Panel Menu Reference SCPI Command
Select System\Admin\Cal\Volt\Vlim.
Select the High range. Then press Next.
CAL:VOLT:LIM:POS 20, (@1)
Step 2. Select the first voltage calibration point and enter the data.
Front Panel Menu Reference SCPI Command
Display shows: “Enter P1 measured data” Enter the data from the external
DMM. Press Select when done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
6 Verification and Calibration
362 Keysight N6700C Operating and Service Guide