Test Record Form - Keysight N6781A, N6782A, and N6784A
N6781A, '82A, and '84A Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Constant Voltage Tests
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
All
All
All
All
1.8 mV
Vout 1.2 mV
19.9932 V
Vout 6.2 mV
_______
_______
_______
_______
+ 1.8 mV
Vout + 1.2 mV
20.0068 V
Vout + 6.2 mV
Voltage Prog. & Readback Low Range
1 V Voltage Vout:
Voltage measured over interface:
All
All
0.99915 V
Vout 325 μV
_______
_______
1.00085 V
Vout + 325 μV
600 mV Prog. & 100 mV Readback
100 mV Voltage Vout:
Voltage measured over interface:
All
All
99.775 mV
Vout 75 μV
_______
_______
100.225 mV
Vout + 75 μV
CV Load Effect
20 V output range:
6 V output range:
All
All
700 μV
400 μV
_______
_______
+ 700 μV
+ 400 μV
CV Ripple and Noise
peak-to-peak 20 V output range:
rms 20V output range:
peak-to-peak 6 V output range:
rms 6V output range:
All
All
All
All
N/A
N/A
N/A
N/A
_______
_______
_______
_______
+ 12 mV
+ 1.2 mV
+ 12 mV
+ 1.2 mV
Transient response
6 V output range @ 35 μs:
20 V output range @ 35 μs:
All
All
20 mV
10 mV
_______
_______
+ 20 mV
+ 10 mV
Auxiliary voltage readback: N6781A Vout 10 mV _______ Vout + 10 mV
Constant Current Tests
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
All
All
All
All
300 µA
Iout 250 µA
2.9985 A
Iout 1.15 mA
_______
_______
_______
_______
+ 300 µA
Iout + 250 µA
3.0015 A
Iout + 1.15 mA
300mA, 100mA, 10mA Programming
300 mA range Current Iout:
100 mA range Current Iout:
10 mA range Current Iout:
N6781A/N6782A
N6784A
N6784A
99.82 mA
99.958 mA
9.9925 mA
_______
_______
_______
100.18 mA
100.042 mA
10.0075 mA
100 mA, 1 mA, 10 μA Readback
100mA rng measured over interface:
1 mA range measured over interface:
(when installed in N6705 analyzer):
10μA range measured over interface:
(when installed in N6705 analyzer):
All
All
All
Iout 35 µA
Iout 350 µA
(Iout 360 µA)
Iout 10.5 nA
(Iout 22.5 nA)
_______
_______
(_______)
_______
(_______)
Iout + 35 µA
Iout + 350 µA
(Iout + 360 µA)
Iout + 10.5 nA
(Iout + 22.5 nA)
Keysight N6700C Operating and Service Guide 395