3 VMA Mode
3.11 Phase and Amplitude vs Time Measurement
Configuration 2: Mechanical attenuator, no optional electronic attenuator
Note that Configuration2 is not strictly speaking a dual-section attenuator, since
there is no electronic section available. However, it behaves exactly like
Configuration 1 without the Electronic Attenuator Option EA3, therefore for the sake
of this document it is grouped into the “Dual-Attenuator” configuration.
Single-Attenuator Configuration
You can tell which attenuator configuration you have by pressing the Attenuation
tab, which (in most Modes) opens the Attenuation menu. If the first control in the
Attenuation menu says Mech Atten you have the Dual-Attenuator configuration. If
the first control says Atten you have the Single-Attenuator configuration.
(Note that depending on the measurement, there may be no Auto/Man functionality
on the Mech Atten control.)
In the Single-Attenuator configuration, you control the attenuation with a single
control, as the fixed stage has only two states. In the Dual-Attenuator configuration,
both stages have significant range, so you are given separate control of the
mechanical and electronic attenuator stages.
Vector Modulation Analyzer Mode User's &Programmer's Reference 1764