EasyManua.ls Logo

Keysight N9000B - Ref Position; Auto Scaling

Keysight N9000B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 VMA Mode
3.3 Monitor Spectrum Measurement
Ref Position
Positions the reference level at the top, center, or bottom of the Y Scale display.
Changing the reference position does not change the reference level value.
Remote
Command
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:RPOSition TOP | CENTer | BOTTom
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:RPOSition?
Example
:DISP:MON:WIND:TRAC:Y:RPOS CENT
:DISP:MON:WIND:TRAC:Y:RPOS?
Preset
TOP
State Saved Saved in instrument state
Range
TOP|CENTer|BOTTom
Annotation The greater than (>) and less than (<) symbols are displayed on both sides of the graph to indicate the
Reference Position
Backwards
Compatibility
SCPI
:DISPlay:MONitor:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RPOSition
Auto Scaling
Toggles the Auto Scaling function On or Off.
Remote
Command
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:COUPle 0 | 1 | OFF | ON
:DISPlay:MONitor:WINDow[1]:TRACe:Y[:SCALe]:COUPle?
Example
:DISP:MON:WIND:TRAC:Y:COUP OFF
:DISP:MON:WIND:TRAC:Y:COUP?
Couplings When Auto Scaling is ON, and the Restart front-panel key is pressed, this function automatically sets
the scale per division to 10 dB and determines the reference values based on the measurement results
When you change a value of Scale/Div, Ref Value, or Scale Range, Auto Scaling automatically changes
to OFF
Preset 1
State Saved Saved in instrument state
Range
OFF|ON
Backwards
Compatibility
SCPI
:DISPlay:MONitor:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:COUPle
Vector Modulation Analyzer Mode User's &Programmer's Reference 566

Table of Contents

Other manuals for Keysight N9000B

Related product manuals