4 System
4.9 Diagnostics
4.9 Diagnostics
Displays a slider that allows you to view Hardware Statistics.
Dependencies Not available in UXM
4.9.1 Show Hardware Statistics
Provides a display of various hardware statistics. The statistics include the following:
Mechanical relay cycles (on models with mechanical relays)
High and Low temperature extremes
Elapsed time that the instrument has been powered-on (odometer)
Modular instruments display only time and temperature information.
Example
:SYST:SHOW HWST
Notes The values displayed on the screen are only updated upon entry to the screen and not updated while
the screen is being displayed
4.9.2 Pathwave Calibration Advisor…
This is a separate application that helps maintain your instrument at peak
performance. You can set the cal interval, configure cal due reminders, check the
cal status, view cal certificates and test reports, and contact Keysight for a cal
service.
The embedded help documentation can be accessed in the instrument at:
C:\Program Files\Keysight\Calibration Advisor\PCA.chm, or via the ? button at the
top right of the PathWave Calibration Advisor window.
4.9.3 Query the Mechanical Relay Cycle Count (Remote Query Only)
Returns the count of mechanical relay cycles.
Remote Command
:SYSTem:MRELay:COUNt?
Example
:SYST:MREL:COUN?
Notes Query Only
The return value is a comma-separated list of the individual counts for each mechanical relay
The position of the relays in the list is: