EasyManua.ls Logo

Keysight N9000B - 4.9.4 Query the Operating Temperature Extremes (Remote Query Only); 4.9.5 Query the Elapsed Time since 1 st power on (Remote Query Only)

Keysight N9000B
2650 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 System
4.9 Diagnostics
“<Cal Signal>,<AC/DC>,<2dB #1 Atten>,<2dB #2 Atten>,<6dB Atten>,<10dB
Atten>,<20dB Atten>,<30dB Atten>,<Fixed Atten>,<Low Noise Path
Switch>,<Presel Bypass>”
Items in the list not pertaining to your hardware configuration return as -999 for those items
Dependencies Not supported by E6607C
4.9.4 Query the Operating Temperature Extremes (Remote Query Only)
Returns the low operating temperature extreme value. The value survives a power-
cycle and is the temperature extreme encountered since the value was reset by the
factory or service center.
Remote
Command
:SYSTem:TEMPerature:LEXTreme?
Example
:SYST:TEMP:LEXT?
Notes Value is in degrees Celsius at which the lowest operating temperature has been recorded since 1st
power-up
State Saved No
Returns the high operating temperature extreme value. The value survives a power-
cycle and is the temperature extreme encountered since the value was reset by the
factory or service center.
Remote
Command
:SYSTem:TEMPerature:HEXTreme?
Example
:SYST:TEMP:HEXT?
Notes Value is in degrees Celsius at which the highest operating temperature has been recorded since 1st
power-up
State Saved No
4.9.5 Query the Elapsed Time since 1
st
power on (Remote Query Only)
Returns the elapsed on-time in minutes since 1
st
power-on.
Remote Command
:SYSTem:PON:ETIMe?
Example
:SYST:PON:ETIM?
Notes Query Only
1983 Vector Modulation Analyzer Mode User's &Programmer's Reference