3 VMA Mode
3.1 Digital Demod Measurement
“Full Bypass Enabled, maximum safe input power reduced”
Remote Command
[:SENSe]:POWer[:RF]:MW:PATH:AUTO:FULL ON | OFF | 1 | 0
[:SENSe]:POWer[:RF]:MW:PATH:AUTO:FULL?
Example
:POW:MW:PATH:AUTO:FULL ON
:POW:MW:PATH:AUTO:FULL?
Dependencies Only appears if Option FBP is installed, and in the following measurements
5GNRMode: Modulation Analysis and IQ Waveform
WLAN Mode: IQ Waveform
Channel Quality Mode: Group Delay and Noise Power Ratio
Preset
OFF
State Saved Saved in instrument state
Software Preselection
Provided in some instruments, either to compensate for issues with provided
hardware preselection or to provide the preselection function when there is no
hardware preselector.
N9041B
Software Preselection compensates for the frequency range limit of the microwave
preselector. Since the microwave preselector only goes up to 50GHz, software
preselection must be used to suppress and separate images above 50GHz. The
specific algorithm used for software preselection is specified by the SW Preselection
Type selection Normal or Advanced.
In N9041B, Software Preselection only applies for frequencies above 50GHz,
therefore it is only used for RF Input 2. Even if turned on, it is not used for other
inputs, and never for frequencies below 50GHz. This is why the label of this control
contains the parenthetical note “(>~50GHz)”.
Note that in N9041B, in Swept SA measurement, Software Preselection works even
if the measurement is using an FFT Sweep Type. In measurements other than Swept
SA, Software Preselection is not used if the measurement is using an FFT sweep
type.
N9042B+V3050A
Software Preselection compensates for the frequency range limit of the microwave
preselector. Since the microwave preselector only goes up to 50GHz, software
preselection must be used to suppress and separate images above 50GHz. The