6 Input/Output
6.2 Input
11970
Lets you preset for a model in the HP/Agilent/Keysight 11970 series.
Because the X-Series has an LO range of 3.8 - 14GHz, and older analyzers had an
LO range of 3.0 - 6.8GHz, the harmonic numbers used in the X-Series may differ
from those used on older analyzers for the same mixers. Additionally, some of the
11970 mixers cannot be operated over their full range with the X-Series without
switching harmonics. Consequently, you will find that some of the bands (A-Band,
for example) are broken into two ranges for use with the X-Series.
Vector Modulation Analyzer Mode User's &Programmer's Reference 2154