Characteristics and Specifications 5
Electrical Specifications
U1271A/U1272A User’s Guide 159
Resistance
30 Ω 0.001 Ω - 0.2% + 10 0.65 mA - -
300 Ω
[4]
0.01 Ω 0.2% + 5 0.2% + 5 0.65 mA - -
3 kΩ
[4]
0.0001 kΩ 0.2% + 5 0.2% + 5 65 μA- -
30 kΩ 0.001 kΩ 0.2% + 5 0.2% + 5 6.5 μA- -
300 kΩ 0.01 kΩ 0.5% + 5 0.2% + 5 0.65 μA- -
3 MΩ 0.0001 MΩ 0.6% + 5 0.6% + 5
93 nA//
10 MΩ
--
30 MΩ
[5]
0.001 MΩ 1.2% + 5 1.2% + 5
93 nA//
10 MΩ
--
100 MΩ
[5]
0.01 MΩ 2.0% + 10 -
93 nA//
10 MΩ
--
300 MΩ
[6]
0.01 MΩ -
2.0% +10
@ <100 MΩ
8.0% +10
@ >100 MΩ
93 nA//
10 MΩ
--
300 nS
[7]
0.01 nS 1% + 10 1% + 10
93 nA//
10 MΩ
--
Notes for resistance specifications:
1 Overload protection: 1000 Vrms for short circuits with <0.3 A current.
2 Maximum open voltage is <+3.3 V
3 Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent
measurements longer than 1 ms.
4 The accuracy of the 30 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and
thermal effect (by shorting the test leads).
5 For the ranges of 30 MΩ and 100 MΩ, the RH is specified for <60%.
6 The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from –20 °C to 18 °C or
28 °C to 55 °C)
7 The accuracy for ranges <50 nS is specified after the Null function is used on an open test lead.
Table 5-1 DC specifications (continued)
Function Range Resolution
Accuracy Test current
Burden
voltage
Input
impedance
U1271A U1272A (where applicable)