EasyManua.ls Logo

Kikusui PAS 10-35 - Page 112

Kikusui PAS 10-35
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-14 Maintenance PAS SERIES
If you make a mistake in the calibration procedure or connec-
tion and the POWER switch immediately turns OFF when
the OUTPUT
is turned on by performing steps 8 through 13,
the OVP calibration value may be off by a great extent.
In such case, repeat steps 3 through 8 or 9 through 13 until
the PO
WER switch
turns off after 30 s to 60 s. (You may
have to repeat the steps tens of times.)
You are done calibrating the OVP.
Current calibration procedure
Connecting the equipment
1. Turn off the POWER switch.
2. Connect
a shunt resistor and a DVM to the output terminal.
Shown in Fig.7-16.
Fig.7-16Connection for current system calibration
W
arming up
3. Turn on the POWER switch while pr
essing t
he SET switch.
4. Check that the voltage display shows "CAL."
If it is not, press the CURRENT switch.
5. Warm up the equipment including the shunt resistor and
DVM for sufficient time.
NOTE
DVM
HI
LO
Shunt resistor

Table of Contents

Other manuals for Kikusui PAS 10-35

Related product manuals