TOS5200 43
Setting Test Conditions
5
Withstanding Voltage Test
1
Press MORE to display the More Test Conditions setup (More) screen.
2
Press MEMORY 2 or MEMORY 3 to move to the item that you want to set,
and then use the rotary knob to set the value.
Press STOP, or press a key other than MORE to exit the More Test Conditions setup screen
and return to the basic setup screen.
Frequency
Sets the test voltage frequency.
Start voltage (Start Voltage)
Separate from the test voltage, you can set the start voltage of withstanding voltage tests.
You can set the start voltage to 50 % of the test voltage. The output voltage reaches this
value approximately 0.1 seconds after the start of testing.
Voltage fall time (Fall Time)
Separate from the test time, you can set the time that the TOS5200 takes to lower the voltage
after a withstanding voltage test ends. This setting is used only when a PASS judgment
occurs.
Measurement mode (Measurement)
Selects the voltage/current measurement mode.
Memo
On the More screen, use
the MEMORY 2 and
MEMORY 3 keys to move
between settings. Each
time you press one of
these keys, the cursor
moves to the next setting.
After selecting a setting,
use the rotary knob to set
its value.
50 Hz (0) The test voltage frequency is set to 50 Hz (factory default setting).
60 Hz (1) The test voltage frequency is set to 60 Hz.
off (0) The start voltage is not set (factory default setting).
on (1) The start voltage is set to 50 % of the test voltage.
READY LED lit Off
Output on
Output Off
Start voltage
50 %
100 %
0.1 s Voltage rise time
TEST LED lit
TEST LED
blinking
off (0) The output voltage is shut off immediately after a test ends with a PASS
judgment (factory default setting).
on (1) The output voltage falls over approximately 0.1 seconds after a test ends
with a PASS judgment.
RMS (0) Measurement is performed using true rms response (factory default setting).
AVE (1) Measurement is performed using mean-value response.
The measurement mode on Kikusui TOS9200 series, TOS8870A, TOS8850, TOS8850A,
TOS5000 series, and TOS5000A series is mean-value response.