INSTALLATION INSTRUCTIONS Installation
10MAR06
II4825-1
Page
53 of 72
9 Check that the configuration of the “Destination” is correct.
a At the “Service Application” window, select “Diagnostics/Network”.
b Select:
• “Ping”
• “Dicom Echo”
• “Run”
c Type the information for the new “Destination”:
• “IP address”
• “AE Title”
• “Port Number”
d Touch [OK].
e If “Passed” displays, advance to Step g.
f If “Passed does not display, check:
• “Destinations” are correct
• customer network is operating
• STORAGE DEVICE is energized and is set to receive information
g Select “Cancel” from the “Network Diagnostics” window.
10 If all “Destinations” were set, from the “Service Applications” window touch:
• [Connect]
• [Save Configuration]
11 At the prompt, store the data to the MIM DISK.
12 Close the “Service Applications” window.
13 Double-click the “CR Shutdown” icon.
14 Energize the CR 975 SYSTEM.
15 From the main menu touch [Key Operator Functions].