5.2 Various problems and countermeasures
107
Chapter 5
Status Error description Corrective actions
The DR Detector will not go into
the wall stand or table.
DR Detector is warped.
If the DR Detector is still warped even
after the detector is placed on a smooth
surface, contact Konica Minolta technical
representatives.
For AeroDR 1417HQ, AeroDR 1417S, AeroDR
1717HQ, AeroDR 1012HQ, AeroDR 2 1417HQ
and AeroDR 2 1417S
• The protective cover is deformed.
Contact Konica Minolta technical
representatives.
For AeroDR 3 1417HD, AeroDR 3 1717HD
and AeroDR 3 1012HQ
• The back board is deformed.
The DR Detector cannot be
inserted in the AeroDR Battery
Charger
*1
or AeroDR Battery
Charger2.
For AeroDR 1417HQ, AeroDR 1417S, AeroDR
1717HQ, AeroDR 1012HQ, AeroDR 2 1417HQ
and AeroDR 2 1417S
• The protective cover is deformed.
Contact Konica Minolta technical
representatives.
For AeroDR 3 1417HD, AeroDR 3 1717HD
and AeroDR 3 1012HQ
• The back board is deformed.
Foreign material is in the wired connection
connector of the DR Detector.
Refer to "6.1.2 Cleaning" and remove the
foreign material.
Foreign material is in the bottom of the
AeroDR Battery Charger or AeroDR Battery
Charger2.
Contact Konica Minolta technical
representatives.
The I/F Cable cannot be
connected to the DR Detector.
Part of the wired connection connector of the
DR Detector is deformed.
Contact Konica Minolta technical
representatives.
The spring connector of the I/F Cable is de-
formed.
Foreign material is in the wired connection
connector of the DR Detector.
Refer to "6.1.2 Cleaning" and remove the
foreign material.
Foreign material is in the spring connector on
the I/F Cable.
Only wired connection with the
DR Detector cannot be used.
The wired cable is not connected properly.
Check that the wired cable is properly con-
nected to the DR Detector.
Only wireless connection with the
DR Detector cannot be used.
There is an error in the access point.
Check that the Ethernet cable is properly con-
nected to the access point.
The DR Detector and access point are being
used under poor conditions.
• Wireless does not connect
• Wireless communication terminates
• Cycle time is extended
Check the installation location of DR Detector
and access point.
When performance is noticeably lower than at
the initial installation of this device, it is pos-
sible that the installation environment or usage
environment have changed.
DR Detector wired connection
connector is hearted up.
DR Detector wired connection connector is
heated up immediately after charging with
AeroDR Battery Charger
*1
or AeroDR Battery
Charger2.
This is caused by charging and is not a mal-
function.
Charging sometimes takes lon-
ger.
Charging may take some time if the battery is
completely discharged.
It takes time for internal components to start.
Because it is not abnormal, wait a bit.
Battery LED (blue or orange)
ashes fast.
Usage time with the battery has gotten shorter.
It could be that the charging function of the
battery has deteriorated. It can be replaced
with a new battery for a fee if necessary.
The number of images that can be exposed has decreased.
The charging time has gotten shorter.