Section 5 Reconfiguring the Unit for Operation
with a Canopy Connection
Note: After canopy installation, you MUST reconfigure the Logic+ to have a
canopy.
1. Press the
Menu
button.
2. Press the button until the
Services
option is highlighted
(it will turn yellow).
3. Press
OK/Mute
to enter the Services menu screen.
4. Press the button until the
Configuration
option is
highlighted (it will turn white).
5. Press
OK/Mute
to alter the configuration.
6. You will see a
warning screen
alerting you that you are
about to alter the BSC’s settings.
7. Press
OK/Mute
to input the password.
8. When requested for the password press
Light
,
UV light
,
Timer
,
Timer
then
OK/Mute
.
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Calibration
Configuration

Service
Configuration
PASSWORD:
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Configuration
OK to proceed, MENU to return