Quickstart
How do I run a diagnostic on the keypad, relays and contacts?
1. Press the
Menu
button.
2. Press the button until the
Services
option is highlighted (it will
turn yellow).
3. Press
OK/Mute
to enter the Services menu screen.
4. Press the button until the Diagnostic
Diagnostic
option is
highlighted (it will turn white).
5. Press
OK/Mute
to enter the Diagnostic submenu.
6. Press the button until the Diagnostic
Keypad-Relays
option is
highlighted (it will turn white). Press
OK/Mute
.
7. Press any button on the keypad (except
Menu
), and note the
display.
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or to choose,
OK to select, MENU to escape
or to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics

Service
or to choose,
OK to select, MENU to escape
Keypad -Relays
Sash sensors
Canopy Alarm
Airflow Sensor

Diagnostics

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30241 and is the answer not in the manual?

Labconco Purifier Logic+ 30241 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30241
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals