QuickChart Logic+ Type A1 - Imperial Units
Model
30132
30142
30152
30162
Type
A1
A1
A1
A1
Cabinet Size (in feet)
3
4
5
6
Sash Opening (inches)
12
12
12
12
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
21 (3x7)
24 (3x8)
30(3x10)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.55
5.55
5.55
Distance between columns
4.08
5.21
5.39
5.50
Inflow/Exhaust Data
Nominal Average Inflow (FPM)
85+/-5
85+/-5
85+/-5
85+/-5
Sash Open Area (Sq. Ft)
3.04
4.04
5.04
6.04
Nominal Avg. Exhaust Vol. (CFM)
259
344
428
514
Avg. Exhaust Vol. Range (CFM)
243-274
323-364
403-454
483-544
Secondary Inflow Data
3
Sash Opening Template
1
3+
4+
5+
6+
Sensor distance (inches)
2
4
3.25
3.25
4
# of Test points
6
8
10
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (FPM)
259
344
428
281
Avg. Inflow Vel. Range (FPM)
243-274
323-364
403-454
264-297
Correction Factor (CF)
1
1
1
1.83
Average Inflow Volume Range (AIV)
(Avg. velocity x CF)
243-274
323-364
403-454
483-544
Sash Open Area (Sq. Ft)
3.04
4.04
5.04
6.04
Inflow Velocity Range
(AIV / Sash open area)
85+/-5
85+/-5
85+/-5
85+/-5
HEPA Filter Leak Test Data
Air Displacement (CFM)
600
800
1000
1200
Laskin Nozzles needed
1
2
2
2
Theoretical aerosol conc. (ug/l)
3
23
34
27
23
Actual aerosol conc. (ug/l)
4
12
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
36x18x3.06
48x18x3.06
60x18x3.06
72x18x3.06
Performance (CFM)
335
445
555
665
Performance (Pressure in. H
2
O)
.62+/-.1
.5+/-.1
.5+/-.1
<.42”
Labconco P/N
3838400
3838401
3838402
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
18x18x5.88
26x18x5.88
36x18x5.88
48x18x5.88
Performance (CFM)
266
353
441
529
Performance (Pressure in. H
2
O)
< 0.35”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838500
3838501
3838502
3838503
Motor/Blower Data
Labconco P/N
6
3832200
3832201
3832207
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F25T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
Fluor. Lamp Labconco Part #
9721901
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889200
3889201
3889202
3889203
Canopy Slot Area (F
2
)
.12
.12
.12
.12
Canopy Inflow Range (CFM)
50-100
50-100
50-100
50-100
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal Canopy Slot Volume (CFM)
50
50
50
50
Canopy Vacuum (In. H
2
O)
0.1-0.15
0.1-0.15
0.1-0.15
0.1-0.15
Nominal Total Exhaust Volume -
Cabinet + Canopy (CFM)
7
309
394
478
564
Traverse Total Exhaust Volume -
Cabinet + Canopy (CFM)
8
340
433
526
620

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Purifier Logic+ 30241 and is the answer not in the manual?

Labconco Purifier Logic+ 30241 Specifications

General IconGeneral
BrandLabconco
ModelPurifier Logic+ 30241
CategoryLaboratory Equipment
LanguageEnglish