EasyManua.ls Logo

Labconco Purifier Logic+ 30241 - Page 28

Default Icon
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Service 1-800-522-7658
28
QuickChart Logic+ Type B2 Metric Units
Model
30348
30368
Type
B2
B2
Cabinet Size (in m)
1.22
6
Sash Opening (mm)
203
8
Starting Serial #
1303_
1303_
Downflow Data
Nominal Avg. Downflow (m/s)
.279+/.03
55+/-5
grid # of points (rows x columns)
24 (3x8)
36(3x12)
Grid distance from back & sides
152
6.0
Distance between rows (mm)
146
5.75
Distance between columns (mm)
132
5.50
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
105+/-5
Sash Open Area (Sq. m)
.250
.374
Nominal Avg. Exhaust Vol. (m3/s)
.134
.200
Avg. Inflow Vol. Range (m3/s)
.127-.140
.190-.209
Secondary Inflow Data
3
Sash Opening Template
1
4+
6+
Sensor distance (mm)
2
83
4
# of Test points
8
12
Test point location
*
*
Nominal Avg. Inflow Vel. (m/s)
1.44
1.17
Avg. Inflow Vel. Range (m/s)
1.37-1.50
1.12-1.23
Correction Factor (CF)
.0929
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.127-.140
.190-.209
Sash Open Area (Sq. m)
.250
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
Supply HEPA Leak Test Data
Air Displacement (m3/s)
.210
.471
Laskin Nozzles needed
1
1
Theoretical aerosol conc. (ug/l)
3
30
20
Actual aerosol conc. (ug/l)
4
15
12
Exhaust HEPA Leak Test Data
Air Displacement (m3/s)
.314
.471
Laskin Nozzles needed
2
2
Theoretical aerosol conc. (ug/l)
3
41
27
Actual aerosol conc. (ug/l)
4
25
19
Exhaust Data
DIM Exhaust Volume (m3/s)
5
.314
.471
Traverse Exhaust Volume (m3/s)
6
.382
.575
Differential Pressure (Pa)
7
374
448
Exhaust (Inflow) Alarm Data
Alarm Setpoint (m3/s Inflow)
.071 +/- .03
.105 +/-.03
Supply HEPA Data
Width x Depth x Height (in.)
8
48x18x3.06
72x18x3.06
Performance (CFM)
445
665
Performance (Pressure in. H
2
O)
.5+/-.1
<.42
Labconco P/N
3838401
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
8
26x18x8.08
48x18x8.08
Performance (CFM)
720
1100
Performance (Pressure in. H
2
O)
.48+/-.07
.37+/-.07
Labconco P/N
3438501
3438503
Motor/Blower Data
Labconco P/N
6
3832209
3832210
Motor HP
½
½
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721903
Fluor. Ballast Labconco Part #
3838100
3838100

Table of Contents