QuickChart Logic+ PuriCare Procedure Stations
Imperial Units
Model
31241
31142
31261
31162
Type
A2
A1
A2
A1
Cabinet Size (in feet)
4
4
6
6
Sash Opening (inches)
10
12
10
12
Starting Serial #
1303_
1303_
1303_
1303_
Downflow Data
Nominal Avg. Downflow (FPM)
55+/-5
55+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides
6.0
6.0
6.0
6.0
Distance between rows
5.55
5.55
5.55
5.55
Distance between columns
5.21
5.21
5.50
5.50
Inflow Data
Nominal Average Inflow (FPM)
105+/-5
85+/-5
105+/-5
85+/-5
Sash Open Area (Sq. Ft)
3.37
4.04
5.03
6.04
Nominal Avg. Exhaust Vol. (CFM)
354
344
529
514
Avg. Exhaust Vol. Range (CFM)
337-371
323-364
503-553
483-544
Secondary Inflow Data
3
Sash Opening Template
1
3.81
3.81
4.95
4.95
Sensor distance (inches)
2
3.25
3.25
4
4
# of Test points
8
8
12
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (FPM)
354
344
289
281
Avg. Inflow Vel. Range (FPM)
337-371
323-364
275-302
264-297
Correction Factor (CF)
1
1
1.83
1.83
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
337-371
323-364
503-553
483-544
Sash Open Area (Sq. Ft)
3.37
4.04
5.03
6.04
Inflow Velocity Range
(AIV / Sash open area)
100-110
85+/-5
100-110
85+/-5
HEPA Filter Leak Test Data
Air Displacement (CFM)
794
794
1189
1189
Laskin Nozzles needed
2
2
2
2
Theoretical aerosol conc. (ug/l)
3
34
34
23
23
Actual aerosol conc. (ug/l)
4
18
18
15
15
Supply HEPA Data
Width x Depth x Height (in.)
5
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H
2
O)
.5+/-.1
.5+/-.1
<.42
<.42
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Width x Depth x Height (in.)
5
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
353
353
529
529
Performance (Pressure in. H
2
O)
< 0.28”
< 0.28”
< 0.28”
< 0.28”
Labconco P/N
3838501
3838501
3838503
3838503
Motor/Blower Data
Labconco P/N
6
3832201
3832201
3832208
3832208
Motor HP
½
½
¾
¾
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8
TL741
F32T8
TL741
F40T8
SP41
F40T8
SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100
Ventus Canopy Data
Labconco Canopy P/N
3889201
3889201
3889203
3889203
Canopy Slot Area (F
2
)
.12
.12
.12
.12
Canopy Inflow Range (CFM)
50-100
50-100
50-100
50-100
Nominal Canopy Slot Velocity (FPM)
240-260
240-260
240-260
240-260
Nominal Canopy Slot Volume (CFM)
50
50
50
50
Canopy Vacuum (In. H
2
O)
0.1-0.15
0.1-0.15
0.1-0.15
0.1-0.15