EasyManua.ls Logo

Labcyte Echo 525 - Diagnostics Window

Labcyte Echo 525
171 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Diagnostics Window
The Diagnostics window provides the following control and monitoring options:
l In the Device control and status section, the user can control many of the Echo internal devices and mechanisms. The
current position of those devices and mechanisms is reported by green lamp indicators. Clicking the check box next to the
device initiates the action (for example, clicking Source Plate In directs the Echo Liquid Handler to retract the source plate
gripper stage).
l In the System Status section, the user can monitor the current overall status of the Echo Liquid Handler through ten
additional status indicators and two internal Echo temperature readouts.
l In the Miscellaneous section, the user can either activate the Echo maintenance mode or execute diagnostic procedures.
Echo Liquid Handler Software CHAPTER 5 | Software
116