7 | Instrument Specifications
Each Echo Liquid Handler conforms to the specifications listed below.
l Communications
l Fluid Transfer
l Labware
l Electrical
l Ratings
l Physical
l Environmental
l Client PC (minimum requirements)
l Optional Hardware
Communications
Server connection: Echo Liquid Handling Platform (Server Version 2.6.x)
Network connection: 10/100/1000 BaseT
Network protocol: TCP/IP
User account: Labcyte (The client PC must have a Labcyte account, with administrator privileges, that is accessible by Labcyte
Service and Support to service the Echo Liquid Handler system.)
Fluid Transfer
Drop transfer resolution: 25 nL
Volume transfer range:
25 nL to 5 µL (one well to one well)
25 nL to 45 µL (one well to many wells)
25 nL to Working Range (one to many wells)
Note: Actual transfer range is dependent on fluid type and assay plate.
Plate Maximum Minimum Working Range
384PP 65 µL 15 µL 50 µL
384PP Plus 65 µL 20 µL 45 µL
384LDVPlus 14 µL 4.5 µL 9.5 µL
6RES 2800 µL 250 µL 2550 µL
Table 15: Fluid Transfer Ranges by Plate Table
Transfer Accuracy: <10% deviation from target volume
Transfer Precision: <8% CV
LABCYTE INC. USER GUIDE|Echo® 525 Liquid Handler User Guide