l House supply: Minimum 200 Torr (266 mbar, 22 in Hg Vac) minimum pumping speed 8.0 m
3
/hr
l Standalone vacuum with surge tank (10 L): 100 Torr (133 mbar, 26 in Hg Vac), minimum pumping speed 2.0 m
3
/hr
Note: A standalone Vacuubrand MZ 2S-NT vacuum system can be purchased from Labcyte (Part #001-
8104).For more information, see the Vacuum Pump topic.
Note: The 10 L surge tank comes standard with the Echo Liquid Handler to ensure consistent vacuum.
l Electrical Sockets Required:
l One mains outlet for Echo Liquid Handler
l One mains outlet for Chiller
l One mains outlet for Client PC
l One mains outlet for monitor
l One mains outlet for Vacuum pump (if not using house vacuum)
l One mains outlet for Compressor (if not using house air)
61 PN | 001-11665
USER GUIDE | Echo® 525 Liquid Handler User Guide Pre-Installation Requirements